7. Ensure the needle is in "Down" position.
8. Select
Maintenance
;
Prime/Clean, Fill & Clean.
9. Select button
OK
. Now the tube disk can be moved freely.
10. Place the adapter with dummy tube on the tube disk. Ensure tubing between adapter and dummy sample
tube is below the tube disk.
11. Rotate the tube disk until the dummy sample tube is centered over the needle.
12. Close the hood and press
Continue.
13. The needle goes up and the fill and clean process is started. (This cycle takes about 90 minutes).
14. When all the pipettes are filled, the needle goes back to the home position.
15. Remove the adapter unit.
Replace needle O-ring
For assembling and disassembling the special needle holder tool TLAX110023 and spanner TLAX110026 are
needed.
Needle removal
Work instruction Interrliner
MRN-164_2-EN Version 1.08 Starrsed TL
99