E251: Timeout while preparing to open
gripper!
Timeout in ‘prepare to open’-routine.
•
Faulty gripper ‘close’-sensor
detection.
•
Faulty internal mechanics of
the gripper.
•
Faulty gripper motor.
•
Faulty TLAX PCB(s).
•
E252: Timeout while opening gripper!
Timeout in ‘open’-routine.
•
Faulty gripper ‘open’-sensor
detection.
•
Faulty internal mechanics of
the gripper.
•
Faulty gripper motor.
•
Faulty TLAX PCB(s).
E253: Timeout while closing gripper!
Timeout in ‘close’-routine.
•
Faulty gripper ‘close’-sensor
detection.
•
Faulty internal mechanics of
the gripper.
•
Faulty gripper motor.
•
Faulty TLAX PCB(s).
E254: Error while preparing to open
gripper!
Error in ‘prepare to open’-routine.
•
Faulty gripper ‘close’-sensor
detection
E255: Error while opening gripper!
Error in ‘open’-routine.
•
Faulty gripper ‘open’-sensor
detection.
•
Faulty internal mechanics of
the gripper.
•
Faulty gripper motor.
•
Faulty TLAX PCB(s).
E256: Error while closing gripper!
Error in ‘close’-routine.
•
Faulty gripper ‘close’-sensor
detection.
•
Faulty internal mechanics of
the gripper.
•
Faulty gripper motor.
•
Faulty TLAX PCB(s).
•
E257: Error while picking up a tube!
Something went wrong while picking
up a tube. Clear this error to see the
specific error prior to this error.
•
Clear error to see specific
error.
Appendix for Interrliner
MRN-164_2-EN Version 1.08 Starrsed TL
131
Summary of Contents for STARRSED TL
Page 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Page 2: ......
Page 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Page 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Page 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...