Fill and clean
1. Fill beaker with hot de-ionized water (± 130 ml, ±80°C).
2. Add ±15 ml cleaning agent (QRR 010905) to the hot de-ionized water.
3. Mix the solution well.
4. Fill adapter TLAX110906 with the cleaning solution.
- Turn adapter upside down and remove red cap (from dummy tube) and start filling the adapter until the
dummy sample tube is filled without air.
- Place red cap on dummy sample tube and turn adapter again, fill the adapter completely. Check for air in the
dummy sample tube.
5. Ensure that the Interrliner is in Service mode, the status is also indicated with orange LED light.
6. Enable "Sample Mode On"
7. Ensure the needle is in "Down" position.
8. Select
Maintenance
;
Prime/Clean, Fill & Clean.
9. Select button
OK
. Now the tube disk can be moved freely.
Work instruction Interrliner
108
MRN-164_2-EN Version 1.08 Starrsed TL