10.2.1 - Check the sensors in service mode
Vacuum pressure check
▪
Go to tab
Maintenance
->
Check sensor
. Select
Check Flow sensor
box.
Flow:
0980
± 60 Abs:
0320
±10
If the flow is not in range there might be a blockage in the vacuum flow line to the flow sensor.
Fill Stop sensor check
▪
Go to tab
Maintenance
->
Check sensor
. Select
Check Fill stop sensor
box.
Fill stop sensor FS 90..
140
..165
Diluter Start sensor check
▪
Go to tab
Maintenance
->
Check sensor
. Select
Diluter start sensor
box.
Diluter start sensor 400-700
Measure sensor check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check measure sensor
box.
Measure sensor MS .
50
±10
Temperature sensor check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check Temperature sensor
box.
Temperature sensor TS [Room temperature]
Diluent flow sensor check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check Diluent flow sensor
box.
Press test. When test is finished, signal Standby and Flow must shown as activated.
Separator check
▪
Go to tab
Maintenance
->
Check sensor.
Select
Check Separator sensor
box.
Separator sensor <200
600
>700
10.2.2 - Blood spill removal
It is possible blood is leaking during aspiration of the tube. When blood remains on the needle drip block, it
will be drained automatically.
In case of blood leakage outside the needle drip block, blood is collected in the drip tray under the
processing unit.
Check on daily base if blood or other fluid is collected in the drip tray.
1. Remove the drip tray.
2. Clean it in case of containment
3. Place the drip tray back in the instrument.
Maintenance
MRN-164_2-EN Version 1.08 Starrsed TL
67
Summary of Contents for STARRSED TL
Page 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Page 2: ......
Page 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Page 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Page 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...