The tubes and the reagents have the following color codes:
Reagent
Color code
Starrsed Rinse solution
Green
Starrsed Diluent
Grey
De-ionized water
Blue
Starrsed Disinfectant
White
5. Check if reagent tubing is not pinched off.
6. Close cabinet.
7. Enter new reagent information, see
After each reagents change, the fluid system must be primed:
1. Select
Maintenance
->
Prime / Clean
.
2. Perform the applicable prime step to fill the relevant tubes with reagent and remove air.
Maintenance
MRN-164_2-EN Version 1.08 Starrsed TL