- Place red cap on dummy sample tube and turn adapter again, fill the adapter completely. Check for air
in the dummy sample tube.
5. Select Interrliner on the track management system and put instrument "offline"
6. Ensure that the Interrliner is in Service mode, the status is also indicated with orange LED light.
7. Enable "Sample Mode On"
8. Ensure the needle is in "Down" position.
9. Select
Maintenance
;
Prime/Clean, Fill & Clean.
10. Select button
OK
. Now the tube disk can be moved freely.
11. Place the adapter with dummy tube on the tube disk. Ensure tubing between adapter and dummy
sample tube is below the tube disk.
12. Rotate the tube disk until the dummy sample tube is centered over the needle.
13. Close the hood and press
Continue.
14. The needle goes up and the fill and clean process is started. (This cycle takes about 90 minutes).
15. When all the pipettes are filled, the needle goes back to the home position.
16. Open the hood and remove the adapter unit.
17. Solve any outstanding errors on the track information system.
18. Close the hood
19. Put the Interrliner back online.
Maintenance
MRN-164_2-EN Version 1.08 Starrsed TL
69