11. Place the adapter with dummy tube on the tube disk. Ensure tubing between adapter and dummy sample
tube is below the tube disk.
12. Rotate the tube disk until the dummy sample tube is centered over the needle.
13. Close the hood and press
Continue.
14. The needle goes up and the fill and clean process is started. (This cycle takes about 90 minutes).
15. When all the pipettes are filled, the needle goes back to the home position.
16. Open the hood and remove the adapter unit.
17. Solve any outstanding errors on the track information system.
18. Close the hood
19. Put the Interrliner back online.
Work instruction Interrliner
90
MRN-164_2-EN Version 1.08 Starrsed TL