13.1 - Appendix - Maintenance schedule
Maintenance performed periodically or after number of samples
Maintenance Schedule Interrliner (Example)
Daily Maintenance
Perform End-of-day wash, Check tubing/diluent syringe, Clean instrument
Weekly Maintenance
Perform End-of-day wash, Check tubing/diluent syringe, Clean instrument, Clean Fill nozzle, Clean Liquid separator, Check sensors
Sample volume: 450 per day (5 working days)
Parts
Total amount/year
Or earlier after number of samples
91.000
Replace Fill nozzle O-ring
QWLV050004
12
Replace bacterial air filter (HEPA)
QWLV040002
12
Replace peristaltic tube assembly (2x)
ESRI090902/903
12
Run Fill and clean
Check/Clean Liquid sensor
12
Replace O-ring inner needle
QWLV050002
12
Replace waste pump tube
QWLV090011
4
Appendix for Interrliner
122
MRN-164_2-EN Version 1.08 Starrsed TL
Lev
el
4
Lev
el
3
Lev
el
2
Lev
el
1
7.500
samples
23.000
samples
46.000
samples
91.000
samples
Summary of Contents for STARRSED TL
Page 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Page 2: ......
Page 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Page 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Page 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...