Please note the light signal at the front near the manual input if a
rack can be placed or removed. The rack is moved in and out during
processing of the sample tubes (The rack signal light is red).
1. Start manual sampling in the software with pressing button
Sample
and then button
Manual input
, the
manual input adapter comes out. (Rack Signal light is green)
2. Place a rack with tubes in the manual input adapter.
3. Press button
Manual input
again, the manual input adapter moves to the internal pick position and the
samples are processed. (Tubes from the track are on hold). If the rack signal light is red the instrument is
busy with another process.
4. Sample tubes which were present in the tube disk before the manual input process from earlier track
feeding will be temporarily transported to the 2 buffer tube holders next te the manual input adapter.
5. After processing the last sample tube the tube is placed back in the rack and the manual input adapter
moves out.
6. If applicable the sample tube(s) in the buffer position(s) will transported to the track after manual input
process is closed.
Remove rack, place a new rack or start with track feeding again. With pressing button
Manual input
the
adapter moves in again.
Sampling results of manual input are always visible in patient history and, depending on LIMS
communication settings, also send to the LIMS.
Operation
MRN-164_2-EN Version 1.08 Starrsed TL
57
Summary of Contents for STARRSED TL
Page 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Page 2: ......
Page 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Page 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Page 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...