Warning
By selecting one of the warning codes, all the samples with this warning code are displayed in the table.
In the header of the table the names of the columns are shown. Double-click the header of any column to
sort the table by this column in ascending order.
5.4 - Reagents screen
When there is a sensor alarm, an alarm indicator is shown in the tab
Reagents
.
The alarm status of the reagent containers and separator are shown in this screen. An empty container is
marked by a flashing red to yellow mark.
When the status screen is active, the reagent audio alarm is switched off.
Reagent information is shown in the little text boxes. To input new reagent information when reagent
container is replaced, click on the appropriate text box.
Note: When the expiry date is exceeded the text box will flash red.
The software checks the reagent level status before starting a new aspiration. If a level alarm is
ON
, it will
not process the new sample. If an alarm comes
ON
during an aspiration, it will finish to aspirate that
sample. Washing dirty pipettes always continues, as to avoid that the samples are left in the pipettes.
Reagents alarm is also set when the expire date of the reagent is exceeded or opened more than three
months. The message Not allowed now! See
Reagents
! appears. Processing of new samples is stopped. A log
is available for all reagents and can be accessed by clicking on
Starrsed TL program
MRN-164_2-EN Version 1.08 Starrsed TL
43
Summary of Contents for STARRSED TL
Page 1: ...Instructions for Use Starrsed TL MRN 164_2 EN Version 1 08...
Page 2: ......
Page 7: ...Index 136 Contents MRN 164_2 EN Version 1 08 Starrsed TL 7...
Page 116: ...Work instruction Interrliner 116 MRN 164_2 EN Version 1 08 Starrsed TL...
Page 120: ...Work instruction Interrliner 120 MRN 164_2 EN Version 1 08 Starrsed TL...