11.7 - Leaking pipettes
1. Check for particles like specks of dirt or hairs in the pipette
valve.
2. If no particles are found, replace the valve tube and valve body.
11.8 - Liquid level sensor not sensing
1. Liquid in the container is not detected. This occurs sometimes with the
De-ionized water
container and is
caused by a very low conductivity.
2. Add one or two drops of
Diluent
to the
De-ionized water
to increase the conductivity.
11.9 - Sampling and results troubleshooting
Results are not sent to LIMS:
If samples are processed and results are visible in patient history, but are not sent to the LIMS:
Check in LIMS settings if the correct protocol is used.
Sampling does not start with use of manual input:
Check in LIMS settings if the correct protocol is used.
Tubes are replaced on track or rack without aspiration:
Barcodes cannot be read (visible as red flags on tube in sample screen)
No ESR is requested from LIMS
Sampling with manual input does not start:
Troubleshooting
80
MRN-164_2-EN Version 1.08 Starrsed TL