Work instruction Number
234
Page
1 of 1
Purpose:
Maintenance level 3
Safety:
Bio Hazard area
Instrument
: Interrliner
Revision:
002, 2019
▪
Clean fill nozzle and exchange O-ring Fill Nozzle
▪
Clean liquid separator
▪
Exchange peristaltic pump tubes
▪
Fill & Clean
▪
Check/Clean liquid sensor
▪
Sensor check
▪
Exchange waste pump tube
▪
Exchange pinch valve
▪
Final cleaning and preparation
▪
Air filter replacement
Inactivate the instrument in the LAS.
Clean fill nozzle and exchange O-ring Fill Nozzle
1.
2.
3.
4.
5.
6. Clean fill nozzle
The use of a toothbrush and detergent is
recommended.
▪
Carefully scrub the fill nozzle inner part.
▪
Use a tissue to dry the fill nozzle.
Work instruction Interrliner
104
MRN-164_2-EN Version 1.08 Starrsed TL