Rosemount Model P330 Pressure Validator
A-2
EQUIPMENT NEEDED
It is essential to use the proper testing equipment for calibration. Table
10-2 lists the equipment necessary to ensure accurate testing results.
TABLE 10-2. Calibration Equipment.
CALIBRATION CHECK
PROCEDURE
The calibration check consists of comparing the validator to a known
standard. When performing the check, it is important to:
• Use repeatable and linear pressure sources
• Allow adequate stabilization time before calibration (a minimum
of one hour)
• Conduct the check in a temperature- and humidity-controlled
environment free from drafts, with the recommended
temperature 21 °C ±2 °C.
NOTE
It is recommended that the calibration check be done at 0, 20, 40, 60,
80, and 100% of Full Scale on an ascending and descending run.
Entering the Calibration
Menu
To enter the
CALIBRATION
menu:
1. Press the
SETUP
key.
2. Use
Ä
to highlight
SETTINGS
.
Press
ENTER
.
3. Use
Ä
to highlight
CALIBRATION
.
Press
ENTER
.
4. Type in the four-digit PIN:
a. Use
¶
and
Ä
to scroll through digits.
b. Use
©
to move to next position.
c. When the PIN is entered, press
ENTER
.
NOTE
The default PIN is 4321.
5. In the
CALIBRATION
menu, use
¶
and
Ä
to highlight the desired
parameter for calibration. Press
ENTER
.
6. Follow the prompts on the display to proceed with the test.
7. Press
E
XIT
to exit the
CALIBRATION
menu.
NOTE
Adjustment should be done only by knowledgeable persons. It is
recommended that adjustment be done only by the manufacturer.
Test
Equipment
Pressure Measurement, Internal and External
Class A Compensated Deadweight Tester
Voltage Ranges (5 V and 50 V)
Voltage Source
Current Inputs and Outputs
Digital Milliammeter
Temperature
Precision Temperature Meter
Summary of Contents for P330
Page 2: ......
Page 4: ...Rosemount Model P330 Pressure Validator ...
Page 8: ...iv ...
Page 12: ...Rosemount Model P330 Pressure Validator 1 4 ...
Page 20: ...Rosemount Model P330 Pressure Validator 2 8 ...
Page 30: ...Rosemount Model P330 Pressure Validator 3 10 ...
Page 40: ...Rosemount Model P330 Pressure Validator 4 10 ...
Page 68: ...Rosemount Model P330 Pressure Validator 6 10 ...
Page 80: ...Rosemount Model P330 Pressure Validator 8 6 ...
Page 91: ... 3 ...