6-7
Hardware Functions
Setting Outputs
1. Ensure that the validator is in Advanced task mode.
2. Press the
OUTPUT
key.
3. Use the
¶
and
Ä
to highlight the desired output. Press
ENTER
.
4. Depending on the Output selected, certain parameters will have
to be set. See Output Parameter Descriptions below.
Output Parameter
Descriptions
mA Step
M
A
STEP
is selected using the Setting Outputs procedure above. Once
selected, you must choose an mA Step mode (Figure 6-11):
On selection of mA Step mode, the screen changes to that shown in
Figure 6-12 (in this example, Linear 4–20 mA):
FIGURE 6-12. Display in Output, mA
Step, Linear 4–20 mA Mode.
Press
F1
to run or stop the test. Press
F2
to turn 24 V power on or off.
mA Ramp
M
A
RAMP
must be defined by start and end values, which are entered as
shown in Figure 6-13:
FIGURE 6-13. Entering Start and End
Values for mA Ramp.
FIGURE 6-11. Selecting mA Step
Output and Setting mA Step Mode.