Rosemount Model P330 Pressure Validator
4-8
Setting units in any task mode other than Basic or Advanced
1. Press the
OUTPUT
key.
2. Use
¶
and
Ä
to highlight
UNITS
. Press
ENTER
.
3. Use
¶
and
Ä
to highlight desired units. Press
ENTER
.
4. If desired units are not listed, refer to Selecting Units on page
4-4.
Setting units in Basic mode
1. Press
F2
.
2. Use
¶
and
Ä
to highlight
UNITS
. Press
ENTER
.
3. Use
¶
and
Ä
to highlight desired units. Press
ENTER
.
4. If desired units are not listed, refer to Selecting Units on page
4-4.
Setting units in Advanced mode
1. Press the
OUTPUT
key.
2. Press
F2
.
3. Use
¶
and
Ä
to highlight
UNITS
. Press
ENTER
.
4. Use
¶
and
Ä
to highlight desired units. Press
ENTER
.
5. If desired units are not listed, see Selecting Units on page 4-4.
Zeroing the Display
Both the input and the output displays should be set to zero before a
validation.
NOTE
If the display is more than 5% of zero and an attempt is made to zero it,
the screen displays the message
ZERO TOO LARGE
, and the display will
not zero. Turn the release valve counterclockwise to make sure the
system is depressurized.
To zero the input and output displays, press the shaded keys in Figure
4-10, and proceed as follows:
FIGURE 4-10. Zeroing the Input and
Output Displays.
P
3
30
_5
4
A
Summary of Contents for P330
Page 2: ......
Page 4: ...Rosemount Model P330 Pressure Validator ...
Page 8: ...iv ...
Page 12: ...Rosemount Model P330 Pressure Validator 1 4 ...
Page 20: ...Rosemount Model P330 Pressure Validator 2 8 ...
Page 30: ...Rosemount Model P330 Pressure Validator 3 10 ...
Page 40: ...Rosemount Model P330 Pressure Validator 4 10 ...
Page 68: ...Rosemount Model P330 Pressure Validator 6 10 ...
Page 80: ...Rosemount Model P330 Pressure Validator 8 6 ...
Page 91: ... 3 ...