6-9
Hardware Functions
NOTE
To conserve battery power, keep 24 V off when not in use.
USER TASK MODE
In User Task mode, the validation test is named by the user and saved
in a memory location under the
USER TASKS
menu. There are ten
memory locations for User Tasks.
To define a new task, refer to Figure 6-18 and proceed as follows:
FIGURE 6-18. Naming a User Task.
1. Press the
INPUT
key. Select the desired input for the task, as
described in Setting Input Parameters on page 6-3.
2. Select any Process functions, if necessary, as described in Setting
Process Functions in the Input Menu on page 6-3.
3. Press the
OUTPUT
key. Select the desired output for the task, as
described in Setting Outputs on page 6-7.
4. Press the
TASK
key. Use
¶
and
Ä
to highlight
USER TASKS
. Press
ENTER
.
5. In the
USER
TASKS
menu, use
¶
and
Ä
to highlight a free space.
6. Press
F1
to name the task, then proceed as follows:
a. Use
¶
and
Ä
to enter digits.
b. Use
§
and
©
to move to next or previous digit.
c. Press
ENTER
when the correct task name is displayed.
7. Press
F1
to save the new task or
F2
to clear the task.
NOTE
F2
can be used to clear any highlighted task.
Upon completion of defining a user task, the display looks like that in
Figure 6-19:
FIGURE 6-19. Display of User-Defined
Task.
P
3
3
0_2
7A
P
3
30_
25
A
Summary of Contents for P330
Page 2: ......
Page 4: ...Rosemount Model P330 Pressure Validator ...
Page 8: ...iv ...
Page 12: ...Rosemount Model P330 Pressure Validator 1 4 ...
Page 20: ...Rosemount Model P330 Pressure Validator 2 8 ...
Page 30: ...Rosemount Model P330 Pressure Validator 3 10 ...
Page 40: ...Rosemount Model P330 Pressure Validator 4 10 ...
Page 68: ...Rosemount Model P330 Pressure Validator 6 10 ...
Page 80: ...Rosemount Model P330 Pressure Validator 8 6 ...
Page 91: ... 3 ...