4-9
Hardware Functions
1. Press the
ZERO
key.
The soft box displays the change to
ZERO INPUT
and
ZERO OUTPUT
or
ZERO INTERNAL SENSOR?
depending on the task mode.
2. To zero the input, press
F1
.
3. Press
F1
again to zero the display, or press
F2
to return to the
task screen without zeroing the display.
4. Press the
ZERO
key.
The soft box displays the change to
ZERO INPUT
and
ZERO OUTPUT
.
5. To zero the output, press
F2
.
6. Press
F1
again to zero the output display, or press
F2
to return to
the task screen without zeroing the display.
Turning the 24 V Power
Source On or Off
If the validation test is validator powered, the 24 V power must be
turned on. If the validation test is externally powered, the 24 V power
should be turned off.
To set the 24 V power in Basic task mode, refer to Figure 4-11 and
proceed as follows:
Setting the 24 V power in Basic task mode
1. If in Basic Task mode, press the
TASK
key.
2. Use
Ä
to highlight any mode other than Basic. Press
ENTER
.
3. Press the
OUTPUT
key.
4. Use
Ä
to highlight
24 VOLT
.
5. Use and
§
to change the setting to
ON
or
OFF
.
6. Press
ENTER
to return to task mode screen.
Setting the 24 V power in any task mode other than Basic
1. Press the
OUTPUT
key.
2. Use
Ä
to highlight
24 VOLT
.
3. Use
§
and
©
to change the setting to
ON
or
OFF
.
4. Press
ENTER
to return to Task mode screen.
FIGURE 4-11. Setting the 24 Volt
Power.
P
3
30_
52A
Summary of Contents for P330
Page 2: ......
Page 4: ...Rosemount Model P330 Pressure Validator ...
Page 8: ...iv ...
Page 12: ...Rosemount Model P330 Pressure Validator 1 4 ...
Page 20: ...Rosemount Model P330 Pressure Validator 2 8 ...
Page 30: ...Rosemount Model P330 Pressure Validator 3 10 ...
Page 40: ...Rosemount Model P330 Pressure Validator 4 10 ...
Page 68: ...Rosemount Model P330 Pressure Validator 6 10 ...
Page 80: ...Rosemount Model P330 Pressure Validator 8 6 ...
Page 91: ... 3 ...