Rosemount Model P330 Pressure Validator
10-2
COMMON PROBLEMS
WITH SOLUTIONS
Use the table below to resolve common problems that may occur when
operating the Model P330.
TABLE 10-1. Solutions to Common
Problems
MAINTENANCE
There are no user-serviceable parts on the validator. The validator must
be serviced only by a Rosemount-approved service provider.
Do not use replacement parts other than those supplied by the
manufacturer or the manufacturer’s agent.
In the event of an instrument malfunction, return the validator to the
manufacturer for repair and warranty work.
To contact a service representative, call the Rosemount service
telephone number on the front page of this manual.
Cleaning
Clean the case and external components only with a damp cloth and
mild detergent.
NOTE
Do not use any type of solvent to clean the validator.
Problem
Action
For More Information:
The pressure reading is not
accurate.
Recalibrate the Model P330
pressure function.
See instructions for
recalibrating the Model
P330 in Appendix A.
No pressure reading is
displayed
Check power source and
connections, especially
pneumatic.
See Section 2: Hardware
Controls and
Connections.
Pressure output reading
does not stabilize, even
with fine tuning.
The instrument may be dirty.
The pneumatic lines may
need to be flushed.
Contact your service
provider. Internal parts
are not accessible for the
Model P330.
Pressure output reading
continously drops.
Plug pressure port to
determine if the leak is
internal
Contact your service
provider. Internal parts
are not accessible for the
Model P330.
Electrical values are not
changing when they should or
are reading zero with no
change.
Check the power source.
Press S
ETUP
, then F2 to get
a reading. Check testing
connections.
Testing connections are
diagrammed and
explained in Section 2:
Hardware Controls and
Connections.
See Safety Messages on page 10-1 for complete warning information.
Summary of Contents for P330
Page 2: ......
Page 4: ...Rosemount Model P330 Pressure Validator ...
Page 8: ...iv ...
Page 12: ...Rosemount Model P330 Pressure Validator 1 4 ...
Page 20: ...Rosemount Model P330 Pressure Validator 2 8 ...
Page 30: ...Rosemount Model P330 Pressure Validator 3 10 ...
Page 40: ...Rosemount Model P330 Pressure Validator 4 10 ...
Page 68: ...Rosemount Model P330 Pressure Validator 6 10 ...
Page 80: ...Rosemount Model P330 Pressure Validator 8 6 ...
Page 91: ... 3 ...