Specifications and Reference
Data
MODEL P330
SPECIFICATIONS
Pressure Specifications
Accuracy
Combined non-linearity, hysteresis, and repeatability are ±0.025% F. S.
for positive or negative pressures.
Pressure Ranges
–14.7 to 10,000 psi (–1 to 690 bar) including gauge and absolute
versions
.
Pressure Port
1
/
8
–27 NPT female.
Temperature Effects
± 0.002% reading per °F averaged over –15 to 105 °F
(–15 to 40.6 °C) and w.r.t. 68 °F (20 °C).
Maximum Working Pressure
The pressure can be exceeded by up to two times its full scale, causing
negligible validation change.
Exceptions:
300 psi (20.7 bar): 1.6 times full scale.
6000 psi (414 bar): 1.2 times full scale.
Burst Pressure
The following table lists the burst pressures for the Model P330
pressure ranges:
Integral Pneumatic Pressure/Vacuum Source
–10 psi to 300 psi (–690 mbar to 21 bar).
No.
Pressure Range
Pneumatic
P330
(1)
(1) A = Absolute, G = Gauge.
Hydraulic
P330
3
4
5
6
7
8
9
0
30 inH
2
O (75 mbar)
150 inH
2
O (374 mbar)
750 inH
2
O (1.9 bar)
100 psi (7 bar)
300 psi (21 bar)
1000 psi (69 bar)
3000 psi (207 bar)
6000 psi (414 b ar)
G
Gor A
G or A
G or A
G or A
G or A
G or A
G or A
No.
Model
Pressure Range
Burst Pressure
3
4
5
6
7
8
9
0
P330 Pneumatic
P330 Pneumatic
P330 Pneumatic
P330 Pneumatic
P330 Pneumatic
P330 Hydraulic
P330 Hydraulic
P330 Hydraulic
30 inH2O
150 inH2O
750 inH2O
100 psi
300 psi
1000 psi
3000 psi
6000 psi
43 psi
43 psi
230 psi
525 psi
525 psi
24000 psi
24000 psi
24000 psi
Summary of Contents for P330
Page 2: ......
Page 4: ...Rosemount Model P330 Pressure Validator ...
Page 8: ...iv ...
Page 12: ...Rosemount Model P330 Pressure Validator 1 4 ...
Page 20: ...Rosemount Model P330 Pressure Validator 2 8 ...
Page 30: ...Rosemount Model P330 Pressure Validator 3 10 ...
Page 40: ...Rosemount Model P330 Pressure Validator 4 10 ...
Page 68: ...Rosemount Model P330 Pressure Validator 6 10 ...
Page 80: ...Rosemount Model P330 Pressure Validator 8 6 ...
Page 91: ... 3 ...