Preparing for Use
R&S
®
FSVA/FSV
48
Quick Start Guide 1321.3066.02 ─ 05
1. Press the SETUP key.
2. Press the "Reference Int/Ext" softkey until it is in the desired state.
External reference signal
It is important that the external reference signal is deactivated when switch-
ing from external to internal reference to avoid interactions with the internal
reference signal.
Remote commands:
ROSC:SOUR EXT
ROSC:EXT:FREQ 20
3.4.3
Setting the Date and Time
You can set the date and time for the internal real time clock as follows:
Opening the Date and Time Properties dialog box
1. Press the SETUP key.
2. Press the "General Setup" softkey.
3. Press the "Time + Date" softkey to open the "Date and Time Properties" dia-
log box.
The "Date & Time" tab is displayed.
Changing the date
1. Press the arrow on the "Month" field to display the list.
2. Select the month from the list.
3. Select the year by clicking on the up and down arrow buttons next to the
"Year" field.
4. Select the day in the calendar display or enter the date via the keyboard.
5. Click "OK".
Remote command
SYST:DATE 2008,10,1