Preparing for Use
R&S
®
FSVA/FSV
37
Quick Start Guide 1321.3066.02 ─ 05
3.1.4
Connecting the AC Power
The R&S
FSVA/FSV is equipped with an AC power supply connector. The
R&S
FSVA/FSV can be used with different AC power voltages and adapts itself
automatically to it. Refer to the datasheet for the requirements of voltage and fre-
quency. The AC power connector is located on the rear panel of the instrument.
► Connect the R&S
FSVA/FSV to the AC power supply using the supplied
power cable.
Since the instrument is assembled in line with the specifications for safety
class EN61010, it may only be connected to an outlet that has a ground con-
tact.
3.1.5
Changing the AC Supply Fuse
Only fuses of the type IEC 60 127-T3.15H/250 V should be used. Those fuses
are used for all of the specified nominal AC supply voltages.
To change the AC supply fuses
1. Disconnect the power cable.
2. Open the flap covering the fuse holder using a small screwdriver (or similar).
3. Remove the fuse holder.
4. Remove both fuses and install the new ones.
5. Reinsert the fuse holder.
6. Close the flap.
Putting into Operation