Brief Introduction to Remote Control
R&S
®
FSVA/FSV
152
Quick Start Guide 1321.3066.02 ─ 05
7.2.1
Default Setting of the R&S
FSVA/FSV
The following settings provide typical examples of how to change the default set-
ting of the R&S
FSVA/FSV.
Note that only some of the settings are necessary depending on the application
example. In many cases, it is not necessary to set resolution bandwidth, video
bandwidth and sweep time since these parameters are automatically calculated in
the default setting when the span is changed. Likewise, the input attenuation is
automatically calculated in the default setting as a function of the reference level.
Last of all, the level detectors are linked to the selected trace mode in the default
setting.
The settings automatically calculated in the default setting are indicated with an
asterisk (*) in the following program example.
7.2.1.1
Setting the Remote Control Status Registers
REM ************************************************************************
Public Sub SetupStatusReg()
'--------- IEEE 488.2 status register --------------------------------------
CALL InstrWrite(analyzer,"*CLS") 'Reset status registers
CALL InstrWrite(analyzer,"*SRE 168") 'Enable Service Request for
'STAT:OPER-,STAT:QUES- and ESR-Register
CALL InstrWrite(analyzer,"*ESE 61") 'Set event enable bit for:
'operation complete 'command-, execution-,
'device dependent- and query error
'--------- SCPI status register --------------------------------------------
CALL InstrWrite(analyzer,"STAT:OPER:ENAB 0")
'Disable OPERation Status reg
CALL InstrWrite(analyzer,"STAT:QUES:ENAB 0")
'Disable questionable Statusreg
End Sub
REM ************************************************************************
7.2.1.2
Default Settings for Measurements
REM ************************************************************************
Public Sub SetupInstrument()
'--------- Default setting f the R&S FSV ---------------------------------
CALL SetupStatusReg 'Configure status registers
Detailed Programming Examples