Front and Rear Panel View
R&S
®
FSVA/FSV
23
Quick Start Guide 1321.3066.02 ─ 05
The output of the tracking generator is to be connected to the DUT via a cable
equipped with a male N connector. The female connector is available only with
the tracking generator option (R&S
FSV-B9).
If the output is active, the LED is green.
Risk of damaging the tracking generator
To avoid damaging the tracking generator, make sure that the reverse
power (i.e. input power caused, for example, by reflections in the test setup)
never exceeds 1 W RF power. Do not apply a DC voltage at the output.
For details, refer to the data sheet.
Sensitive DUTs concerning matching
For DUTs with sensitive RF characteristics with regard to matching (VSWR)
at the input, insert a 10 dB attenuator between the DUT and the tracking
generator.
2.1.4.3
Connections for External Mixers (EXT MIXER, Option R&S
FSV-
B21)
External mixers can be connected at the LO OUT/IF IN and IF IN female connec-
tors (option R&S
FSV-B21 for R&S
FSVA/FSV 30 and R&S
FSVA/FSV 40 instru-
ments). Both two-port and three-port mixers can be used. Connect the mixer as
follows:
Use the supplied coaxial cable to feed in the LO signal. If no external mix-
ers are connected to the R&S
FSVA/FSV, cover the two front connectors
LO OUT / IF IN and IF IN with the SMA caps supplied.
Front Panel View