D-3
D. EMC Declarations
Immunity test
IEC 60601-1-2 test
level
Compliance
level
Electromagnetic
environment - guidance
Surge IEC
61000-4-5
±1 kV line(s) to
line(s) ±2 kV line(s)
to earth
±1 kV
line(s) to
line(s) ±2
kV (line(s)
to earth
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage
dips, short
interruptions,
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
< 5% UT (>95%
dip in UT ) for 0.5
cycle 40% UT(60%
dip in UT) for 5
cycles 70% UT
(30% dip in UT) for
25 cycles < 5% UT
(>95% dip in UT)
for 5 s
<5% UT
(>95% dip
in UT) for
0.5 cycle
40% UT
(60% dip
in UT) for
5 cycles
70% UT
(30% dip in
UT) for 25
cycles <5%
UT (>95%
dip in UT)
for 5 s
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the ventilator
requires continued
operation during power
mains interruptions, it is
recommended that the
ventilator be powered
from an uninterruptible
power supply or a battery.
Power
frequency
(50/ 60 Hz)
magnetic
field IEC
61000-4-8
3 A/m
30 A/m
The power frequency
magnetic field should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment.
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...