7-29
7. Alarms
Item Alarm
Priority
Definition
Action needed
59
CO
2
sensor
faulty, please
reinsert the
sensor
Medium
The mainstream
CO
2
module
sensor
connection is
abnormal.
Please reinsert the
sensor.
60
Pressure Limit
Exceeded
Low
Pressure reaches
pressure
high alarm
limit-5cmH
2
O.
Check parameter
settings, including
pressure high
alarm limit setting.
61
Pressure
Limited in
Sigh cycle
Low
After sigh
function is
activated, the
pressure in sigh
cycle reaches
pressure high
alarm limit
-5cmH
2
O.
Check parameter
settings, including
pressure
high alarm
limit setting.
62
Pressure not
reached
Low
Ppeak fails
to reach
the set value.
Check if the tube
is leaky. If yes,
re-connect.
Check parameter
settings.
63
Pressure not
released
High
The pressure
released
valve is fault.
Check expiratory
valve and
breathing circuit.
64
Pressure
Regulation
High
High
The airway
pressure is
sustained at
a high level.
Confirm ventilator
and alarm settings
are appropriate.
65
Technical
Fault (1037)
High
Insp. flow sensor
failure
Contact your service
personnel.
66
Technical
Fault (1038)
High
Insp. pressure
sensor failure
Contact your service
personnel.
67
Technical
Fault (1039)
High
patient pressure
sensor failure
Contact your service
personnel.
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...