6-27
6. Start Ventilation
compensation during exhalation.
3. Select [
Exit
] for the system to initiate STRC. After
STRC has been enabled, if you enter the STRC
interface and then select [
Disable
], the system will
terminate STRC immediately in the ventilation.
When STRC is enabled, Ptrach waveform is displayed
with the Paw waveform.
WARNING
•
STRC may induce autotriggering. If autotriggering
occurs, first check the patient, breathing circuit,
and other possible causes.
NOTE
•
Incorrect tube type or ID setting can endanger the
patient. Make sure to set them properly.
6.7 Leak Compensation
The leakage from the breathing circuit and mask may
cause that the gas volume delivered to the patient’s
lung is lower than the setting value. The leakage also
may cause the false inspiratory trigger or difficult
switching between inspiratory and expiratory.
The ventilator provides automatic leakage
compensation function. The ventilator updates the
amount of leakage at the end of each breathing cycle
according to the difference between the inspired tidal
volume and expired tidal volume, and the amount of
leakage can be used for the calculation of real-time
leakage flow in next breathing cycle.
During the expiration stage, the base flow will be
regulated automatically to compensate the leakage and
maintain the PEEP valve. In order to prevent the false
inspiratory trigger, flow trigger working mechanism
is based on the compensated flow. Maximum leakage
compensation flow is 65 L/min for adult patients, 45
L/min for pediatric patients and 15 L/min for neonate
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...