11-24
11. Specifications
Mode
Parameter
Factory Default Settings
VS
VT
Adult: max(7*mL/kg*IBW, 100)
Pediatric: 7*mL/kg*IBW
O
2
% (HPO)
21%
f
Adult: 10 /min
Pediatric: 20 /min
I-trigger
3
E-cycle
3
EPAP
10 cmH
2
O
Rise
3
P-min
10 cmH
2
O
P-max
20 cmH
2
O
PPS
O
2
% (HPO)
21%
f
Adult: 10 /min
Pediatric: 20 /min
T-insp.
Adult: 2.00 s
Pediatric: 1.00 s
I-trigger
3
E-cycle
3
IPAP
5 cmH
2
O
EPAP
10 cmH
2
O
Rise
3
P-max
20 cmH
2
O
V-max
600 mL
E-max
15 cmH
2
O/L
R-max
4 cmH
2
O/L/s
PPV%
50%
Table 11-13.
O
2
Therapy settings
Parameter
Factory Default Settings
O
2
%
21%
Flow
25 L/min
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...