7-16
7. Alarms
7.12 Alarm Test
The Ventilator performs a self-check during start-up
and continuously during operation. You may want to
run alarm tests to demonstrate the alarm’s operation.
The sum of the mean alarm condition delay plus the
mean alarm signal generation delay is less than 5s.
7.12.1 Low PEEP
1. Remove the expiration valve membrane and install
the expiration valve.
2. After the ventilator system starts up normally,
connect a test lung to the ventilator and start
ventilation.
3. Set PEEP to 5 cmH
2
O. Verify that the [
Low PEEP
]
alarm is activated.
7.12.2 Patient Circuit Occluded
1. Make sure a demonstration lung is connected to
the ventilator and start ventilation in PCV mode
normally.
2. Nip the inspiration tube with the hands, make sure
the monitoring value of VTi is lower than 10 mL.
3. Verify that the Patient Circuit Occluded alarm is
activated after several breathing cycles.
4. Loosen the inspiration tube and verify this alarm is
reset automatically.
7.12.3 High Oxygen
1. Correctly connect the oxygen supply and close the
air intake.
2. Connect a test lung to the ventilator, set FiO
2
to
60%, and start ventilation.
3. Verify that the [
High Oxygen
] alarm is activated.
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...