3-14
3. Installations and Connections
1
2
1
Oxygen supply inlet
2
Oxygen supply hose
Figure 3-7.Connecting the oxygen supply
To connect the high-pressure oxygen supply as follows:
1. Before connecting the oxygen supply hose, check
if the sealing ring at the gas supply connection is
in good condition. If the sealing ring is damaged,
do not use the hose and replace the sealing ring to
prevent leakage.
2. Align and connect the fitting to the inlet of the
high-pressure oxygen supply at the back of the
ventilator.
3. Make sure the gas supply hose is connected to the
gas supply inlet, then tighten the hose nut.
To connect the low-pressure oxygen supply as follows:
1. Depress the metal dome on the low-pressure
oxygen supply connector.
2. Align the oxygen supply hose with and insert it into
the low-pressure oxygen supply connector.
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...