F. Neonatal Ventilation
F-7
Table F-1.Control settings
Parameter
Range
Resolution
Accuracy
VT
2 to 300 mL
1 mL
20 to 2000mL:
±(10mL+10% of
setting)
2 to 20mL:
±(2mL+10% of
setting)
f
1 to 150 /min
1 /min
±1 /min
f-SIMV
1 to 150 /min
1 /min
±1 /min
T-insp.
0.1 to 10s
0.05s
±0.1s or ±10% of
setting, whichever is
greater
PEEP
OFF, 1 to 25
cmH
2
O
1 cmH
2
O
±(2.0 cmH
2
O+ 5% of
setting)
△ P-insp.
3 to 60 cmH
2
O
1 cmH
2
O
±(2.0 cmH
2
O+ 5% of
setting)
△ P-supp.
0 to 45 cmH
2
O
1 cmH
2
O
±(2.0 cmH
2
O + 5%
of setting)
P-high
0 to 45 cmH
2
O
1 cmH
2
O
±(2.0 cmH
2
O + 5%
of setting)
P-low
0 to 25 cmH
2
O
1 cmH
2
O
±(2.0 cmH
2
O+ 5% of
setting)
T-slope
0 to 0.6 s
0.01s
±(0.2s+20% of
setting)
P-trigger
/
F-trigger
0.1 to 5 L/min
0.1 L/min
±(1.0 L/min + 10%
of setting)
ΔP-apnea
3 to 60 cmH
2
O
1 cmH
2
O
±(2.0 cmH
2
O + 5%
of setting)
VT-apnea
2 to 300 mL
1 mL
±(10mL+10% of
setting)
f-apnea
15 to 150 /min
1 /min
±1 /min
Flow
2 to 12 L/min
1 L/min
±(2 L/min+10% of
setting)
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...