8-6
8. Special Functions
within one minute after nebulization or suction,
within one minute after last PV loop test.
•
The PV tool function is not recommended when
there is great leakage or when the patient has
spontaneous breathing. The relevant characteristic
points the PV tool function provides are only for
your reference.
•
If no operation is performed on PV tool screen
within three minutes, the measurement screen exits
automatically.
The PV tool is the method that records a quasi-static
pressure/volume curve showing both the inflation and
the deflation curve to determine the optimal PEEP. The
doctor is able to determine the optimal PEEP for the
patient with the help of this function.
To perform the PV tool as follow:
1. Select the [
Function
] ->[
PV Tool
].
2. Setup the below parameters for PV tool.
P-start
: starting pressure of the static P-V loop.
P-top
: maximum pressure of the static P-V loop can
reach.
End PEEP
: ending pressure of the static P-V loop.
T-pause
: pause time when the static P-V loop reach
the maximum pressure.
Ramp Speed
: the speed of the starting pressure
rising to maximum pressure during the static P-V
loop.
The system acquires T-total parameter value based
on the calculation formula and displays it on the
window.
3. Press the [
Start]
key to start PV tool measurement.
Press the [
Stop
] key during measurement, the
system aborts measurement immediately.
4. After the measurement is completed, you can set
the desired positions of [
Cursor 1
] and [
Cursor 2
].
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...