7-12
7. Alarms
[
Continuous
]: indicates that the nurse call signal
lasts until the alarm ends, i.e. the duration of a nurse
call signal equals to that of the alarm.
4. Select [
Contact Type
].
[
Normally Open
]: to trigger nurse call with normally
open signal.
[
Normally Closed
]: to trigger nurse call with
normally closed signal.
5. Select [
Alarm Level
] and select levels of alarm that
will trigger nurse call signal.
6. Select [
Alarm Type
] and select types of alarm that
will trigger nurse call signal.
If no setting is made for [
Alarm Level
] or [
Alarm
Type
], nurse call signals will not be triggered no
matter what alarm occurs.
XXXXXXXXXX
XXXXX
XXXXX
XXXXX
XXXXX
XXXXX
XXXXX
XXXXXXXXXX
XXXXXXXXXX
XXXXX
XXXXX
XXXXX
Figure 7-5.Nurse call settings
WARNING
•
Do not rely exclusively on the nurse call system
for alarm notification. Remember that the most
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...