10-3
10. Cleaning and Disinfection
Wipe
: wipe with a damp cloth immersed in medium-
or high-efficiency disinfectant and then wipe off the
remaining detergent with a dry lint free cloth.
Immersion
: immerse it in medium- or high-efficiency
disinfectant (alcohol or isopropyl alcohol, etc.) for more
than 30 minutes (recommended time). Then clean with
water and dry completely.
Steam autoclave at 134ºC for 10 to 20 minutes
(recommended time).
Ultraviolet radiation for 30 to 60 minutes
(recommended time).
The table below lists the cleaning and disinfecting
method can be selected for ventilator part.
Table 10-1.The methods of cleaning and disinfecting
Part
Frequency
Cleaning
Disinfection
Wipe
Imme
rsion
Wipe
Imme
rsion
Autoc
laving
Ultrav
iolet
radia
tion
External
ventilator
surface(including
housing, plug-in,
power cord and
gas supply hose).
Each patient √
√
√
Trolley and
support arm
Each patient √
√
√
Touch screen
Each patient √
√
√
Fan dust filter
Every four
weeks/as
necessary
√
√
Air intake dust
filter
Every four
weeks/as
necessary
√
√
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...