9-4
9. Maintenance
Interval
Part/accessory
Procedure
Each patient or as
necessary
Breathing circuit
(including mask,
inspiratory filter,
flow sensor,
nebulizer jar,
expiratory
valve cover and
membrane).
Replace with sterilized or
new single use parts. Run
the flow sensor calibration.
Entire ventilator.
Run the system self-check.
Every day or as
necessary
Breathing circuit
Empty any water from
breathing tubes or water
traps. Inspect parts for
damage. Replace as
necessary.
Ventilator
Clean the external surfaces.
O
2
sensor
Please calibrate the O
2
sensor. Replace CO
2
sensor
if it is damaged.
Every month or as
necessary
Air intake dust
filters and fan
filter
Check for dust and lint. If
needed, clean or replace.
Check every 6 months
and replace every two
years
Batteries
Check the charging and
discharging of the lithium
battery every 6 months
and replace the lithium
battery every two years.
Contact servive providers
for replacement.
Every 6 years or as
necessary
Battery of the
clock module
Replace the battery of the
clock module. Contact
servive providers for
replacement.
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...