5-8
5. System Settings
XXXXXXXXXX
XXXXXXXXXX
XXXXXXXXXX
XXXXXXXXXX
XXXXXXXXXX
XXXXXXXXXX
XXXXXXXXXX
100%
1.25
Figure 5-7.Ventilation settings windows
5.4.5 Set Oxygen Concentration during Suction
1. Select [
Setup
] → [
Maintain
] → [
Vent
] → [
Setting
].
2. Set [
O
2
% during suction
]: Set oxygen enrichment
in accordance with different patient types. After
initiation of oxygen enrichment, the system will
compare “current oxygen concent oxygen
enrichment” with “100 vol.%” and start ventilation
according to the lower of the two values.
5.4.6 Set Oxygen Sensor Monitoring
1. Select [
Setup
] → [
Maintain
] → [
Vent
] → [
Sensor
].
2. Set the [
Monitoring
]: (ON) or (OFF). When the switch
is ON, oxygen concentration of patient’s inhaled gas
can be monitored. You can switched off if oxygen
concentration monitoring function accompanying
the ventilator is not needed. In this case, the [
O
2
Monitoring Off
] prompt message is displayed on
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...