5-9
5. System Settings
the screen. After Oxygen Sensor Monitoring off, the
ventilator will disable relevant alarm messages and
prompt messages.
NOTE
•
The system total response time for oxygen
concentration monitoring is 15s.
•
The time from powering on the ventilator to
reaching the oxygen concentration monitoring
performance specified in this manual is
approximately 30s. The data sample rate of oxygen
concentration is 100Hz.
CAUTION
•
Disabling oxygen concentration monitoring is
allowable. To prevent potential patient injury, it is
suggested not to switch off oxygen concentration
monitoring continuously.
XXXXXXXXXX
XXXXXXXXXX
XXXXXXXXXX
Figure 5-8.Sensor windows
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...