11-12
11. Specifications
Parameter
Range
Resolution
Accuracy
f-total
0~99 /min
1 /min
±5 % of actual
reading or ±1 /
min, whichever is
greater
f-mand.
0~99 /min
1 /min
±5 % of actual
reading or ±1 /
min, whichever is
greater
f-spn.
0~99 /min
1 /min
±5 % of actual
reading or ±1 /
min, whichever is
greater
I:E
9.9: 1 to 1: 99
9.9:1-1:9.9: 0.1
1:10-1:99: 1
±0.1 or ±10 % of
actual reading,
whichever is
greater.
FiO
2
15 to 100 vol.%
1 vol.%
±(2.5vol.% + 2.5%
of actual reading)
Re
0~600 cmH
2
O/
L/s
1 cmH
2
O/L/s
0~50cmH
2
O/(L/s):
±10cmH
2
O/(L/s)
Other range:
50% of the actual
reading
Ri
0~600 cmH
2
O/
L/s
1 cmH2O/L/s
0~50cmH
2
O/(L/s):
±10cmH
2
O/(L/s)
Other range:
50% of the actual
reading
C-dyn.
0~300 mL/
cmH
2
O
≥
10: 1cmH
2
O
<10: 0.1 cmH
2
O
±(3 mL/cmH
2
O +
25% of the actual
reading)
C-stat.
0~300 mL/
cmH
2
O
≥
10: 1cmH
2
O
<10: 0.1 cmH
2
O
±(3 mL/cmH
2
O +
25% of the actual
reading)
RC-exp.
0~10.0 s
0.01 s
± (0.2 s + 20 % of
actual reading)
Summary of Contents for iHope RS Series
Page 1: ...iHope Ventilator User Manual RS Series...
Page 10: ......
Page 12: ......
Page 26: ...1 14 1 Safety Information For your note...
Page 28: ......
Page 36: ...2 8 2 General Information For your note...
Page 38: ......
Page 58: ...3 20 3 Installations and Connections For your note...
Page 60: ......
Page 74: ......
Page 88: ...5 14 5 System Settings For your note...
Page 122: ...6 32 6 Start Ventilation For your note...
Page 158: ......
Page 168: ......
Page 186: ......
Page 200: ...10 14 10 Cleaning and Disinfection For your note...
Page 202: ......
Page 203: ...11 1 11 Specifications 11 Specifications 11 1 Physical Characteristic...
Page 230: ...11 28 11 Specifications For your note...
Page 232: ......
Page 233: ...A 1 A Pneumatic Diagram A Pneumatic Diagram Figure A 1 Pneumatic diagram...
Page 240: ...C 2 C Communications Interface For your note...
Page 248: ...E 2 E Compliance For your note...