Chapter 5
Component Removal/Installation
1062788
BiPAP Vision Service Manual © Respironics, Inc.
5-31
5.21 Alarm
Follow these steps to remove the alarm (Figure 5-29). Reverse to install.
1. Remove the front panel enclosure (section 5.7).
2. Disconnect the alarm cable from the display control (DC) board
connector J12.
3. Remove the 2 screws that attach the alarm to the bottom enclosure.
4. Lift the alarm from the bottom enclosure.
Figure 5-29: Removing the Alarm
Bottom enclosure
Alarm cable connects
to DC J12
Alarm
6-32 x 1/4 screw (x2)
(2 lbf-in.)