Chapter 6
6-12
BiPAP Vision Service Manual © Respironics, Inc.
1062788
Troubleshooting
Table 6-3 describes
Check
Vent
error codes, including possible causes and
corrective actions.
Table 6-3: Troubleshooting
Check Vent
Error Codes
Error code
Description
Corrective action
100
Real-time clock (RTC) error on the MC.
Replace MC board.
101
MC nonvolatile RAM (NVRAM) cyclic
redundancy (CRC) error
Replace MC board.
102
MC backup battery error.
1. Replace lithium battery.
2. Replace MC board.
103
MC reference voltage (Vref) error
Replace MC board.
200
DC voltage error.
Replace MC board.
201 with
Vent
Inop
The ventilator continues operating when both a
Vent Inop
and a
Check Vent
(error code 201)
occur.
Perform service described in
Section 6.7.
201
DC audible alarm error.
Replace DC board.
202
DC
Check Vent
indicator current error.
Replace DC board.
203
DC
Vent Inop
indicator current error.
Replace DC board.
204
DC backlight voltage error.
Replace DC board.
205
Low DC alarm battery voltage.
1. Recharge battery (Chapter 4).
2. Replace DC board.
206
Touch pad error.
1. Recharge touch pad.
2. Replace DC board.
300
Circulation fan error.
Replace circulation fan.
301
Invalid PC calibration data.
1. Calibrate blower valve.
2. Replace PC board.
303
Blower speed exceeds 16500 revolutions per
minute (RPM) limit.
1. Replace blower.
2. Replace PC board.
306
Oxygen module (OM) disconnected from PC
board.
1. Replace OM cable.
2. Replace OM.
3. Replace PC board.