BiPAP Vision Service Manual © Respironics, Inc.
8-1
Chapter 8. Options and Upgrades
This chapter describes how to install ventilator options and summarizes
upgrade information.
WARNING:
Electrical shock hazard: disconnect the electrical supply before attempting to
make any repairs to the device.
CAUTION:
Electrical components used in this device are subject to damage from
static electricity. Perform all repairs in an antistatic, ESD-protected
environment.
8.1 Options
This section describes how to install these options:
•
Section 8.1.1: proportional assist ventilation/timed (PAV/T) mode.
•
Section 8.1.2: oxygen baffle, an enhancement for ventilator s/n
105999 and below.
8.1.1
PAV/T Mode
Required equipment:
•
Phillips screwdriver
•
Small, long-shaft flat-blade screwdriver
•
EPROM extraction tool
The PAV/T option kit includes:
•
EPROM for DC board
•
ERPOM for PC board
•
EPROM for MC board
•
Alarm B
PAL (MC board)
CAUTION:
To ensure proper ventilator operation, the EPROMs (DC, PC, and
MC
)
must
have the same revision.
1. Remove the DC board from the ventilator. It is possible to remove the
retaining screws and pull the DC board from the enclosure while
leaving the touch pad and LCD connections attached.