Chapter 6
6-16
BiPAP Vision Service Manual © Respironics, Inc.
1062788
Troubleshooting
N/A (MC)
F04 (DC)
1704 (PC)
MC timeout from slave: MC did not request
data from slave within 15 msec.
1. Replace ICB cable.
2. Replace DC board.
3. Replace MC board.
4. Replace PC board.
705 (MC)
E05 (DC)
1705 (PC)
Packet does not fit in message: buffer in
Move
Packet to Message
is full.
Replace subsystem with lit error LED.
706 (MC)
E06 (DC)
1706 (PC)
Invalid timer value: timers not initialized or
invalid data on timer parameter.
Replace subsystem with lit error LED.
707 (MC)
F07 (DC)
1707 (PC)
Invalid delay value: timers not initialized or
requested delay is greater than 1 minute
(prevents rollover).
Replace subsystem with lit error LED.
708 (MC)
F08 (DC)
1708 (PC)
Invalid delay (usec): timers not initialized or
requested delay is greater than 85 usec (use
Delay
for longer delays).
Replace subsystem with lit error LED.
709 (MC)
F09 (DC)
1709 (PC)
General purpose timer (GPT) not initialized
before
Initiate Timers
call.
Replace subsystem with lit error LED.
70A (MC)
F0A (DC)
170A (PC)
Invalid usec ticks: timers not initialized or
PCsed
is too large (>62499 prevents rollover).
Replace subsystem with lit error LED.
70B (MC)
F0B (DC)
170B (PC)
Invalid priority level:
PCsed
priority level value
is invalid.
Replace subsystem with lit error LED.
70C (MC)
N/A (DC)
N/A (PC)
CRC error on third send to PC: calculated value
does not match sent value on third send.
1. Replace ICB cable.
2. Replace PC board.
3. Replace DC board.
4. Replace MC board.
70D (MC)
N/A (DC)
N/A (PC)
NAK received on third send to PC.
1. Replace ICB cable.
2. Replace PC board.
3. Replace DC board.
4. Replace MC board.
70E (MC)
N/A (DC)
N/A (PC)
Tack timeout on third send to PC: Tack not
received within 750 usec of last byte sent.
1. Replace ICB cable.
2. Replace PC board.
3. Replace DC board.
4. Replace MC board.
Table 6-4: Troubleshooting
Vent Inop
Error Codes
Error code
Description
Corrective action