Chapter 5
Component Removal/Installation
1062788
BiPAP Vision Service Manual © Respironics, Inc.
5-9
3. Pry the fuse drawers from the power entry module.
Figure 5-8: Replacing the Fuses
4. Remove and replace both fuses.
5. Reinstall the fuse drawers into the power entry module with arrows on
the front of the drawers pointing to the right.
6. Shut the fuse holder door and snap into place.
7. Reinstall the power cord.
5.5.1
Voltage Selection
The voltage selection is originally set at the factory. Follow these steps to
select ventilator voltage (Figure 5-9):
1. Turn the ventilator off and disconnect the power cord from the back of
the device and the wall outlet.
2. Gently pry open the top of the fuse holder door. The door hinges
downward.
Power entry module
Fuse drawer
(note arrow)
Fuse