Chapter 7
7-34
BiPAP Vision Service Manual © Respironics, Inc.
1062788
Testing and Calibration
BiPAP Vision System Final Test Form
Ventilator serial no.
Service notification
(Respironics use only)
Model no.
Total operating time
Power Indicator and LCD Controls Test
Power indicator lights
Pass
Fail
Brightness control function
Pass
Fail
Contrast control function
Pass
Fail
Pressure Accuracy Test
CPAP setting
Set
Outlet
Unit
Patient
Manometer
Specification
Pass/Fail
5 cmH
2
O
3-7 cmH
2
O
Pass
Fail
10 cmH
2
O
8-10 cmH
2
O
Pass
Fail
20 cmH
2
O
18-22 cmH
2
O
Pass
Fail
Flow Accuracy Test
Flow setting
Flowmeter
Compensated total flow
Pressure
Specification
Pass/Fail
0 LPM
-5 to 5 LPM
Pass
Fail
10 LPM
4.2-15.8 LPM
Pass
Fail
60 LPM
50.2-69.8 LPM
Pass
Fail
120 LPM
105.6-134.4 LPM
Pass
Fail
120 LPM
18-22 cmH
2
O
Pass
Fail
Dynamic Pressure Regulation Test
Exhalation port test complete
Pass
Fail
Flow waveforms created (peak flow ~ 100 LPM, rate ~ 30 BPM)
Pass
Fail
Manometer reading = 18-22 cmH
2
O
Pass
Fail
S/T Mode Performance Test
Vent cycles between low and high pressure
Pass
Fail
IPAP = 33-37 cmH
2
O, EPAP = 3-7 cmH
2
O
Pass
Fail
Rate = 9.3-10.8 BPM
Pass
Fail
Rate reading:
Options and Controls
Invert
,
Bar Graph
, and
Test Alarms
function correctly
Pass
Fail
Record displayed error messages, if any:
Pass
Fail
Record MC board error codes, if any:
Record PC board error codes, if any: