Chapter 5
5-30
BiPAP Vision Service Manual © Respironics, Inc.
1062788
Component Removal/Installation
5.20 Main Control
(MC) Board
Follow these steps to remove the MC board (Figure 5-28). Reverse to install.
1. Remove the PC board (section 5.19).
2. If necessary, remove nylon standoffs from the MC board.
3. Remove the aluminum standoff from the MC board.
4. Remove these cables from the MC board:
•
RS-232 cable from back panel (MC J3-RS 232 port).
•
Nurse call/remote alarm cable (MC J5-remote alarm connector).
•
RS-232 to DC harness (MC J6-DC J5).
5. Remove the 2 screws that attach the MC board from the bottom
enclosure.
6. Lift the MC board from the bottom enclosure.
Figure 5-28: Removing the MC Board
MC board
Tabbed nylon
standoff (x7)
(2 lbf-in.)
Bottom
enclosure
RS-232 cable
(MC J3)
Power connection
from PC board
Screw (x2)
Nurse call/remote
alarm cable (MC J5)
RS-232 to DC cable
(MC J6 to DC J5)
Socketed nylon
standoff (x3)
(2 lbf-in.)
Stainless steel standoff (x1)
(2 lbf-in.)