Chapter 6
6-22
BiPAP Vision Service Manual © Respironics, Inc.
1062788
Troubleshooting
1218
Startup -12 V voltage reference test error:
signal from MUX <-14.04 V or >-10.12 V.
Replace PC board.
1219
Startup -5 V voltage reference test error: signal
from MUX <-5.833 V or >-4.467 V.
Replace PC board.
121A
Startup Vref voltage reference test error: signal
from MUX <3.749 V or >4.445 V.
Replace PC board.
121B
Startup blower voltage reference test error:
signal from MUX >100 mV.
Replace PC board.
121C
Startup ILFR voltage reference test error:
signal from internal ADC >100 mV.
Replace PC board.
121D
Startup oxygen valve voltage reference test
error: OM drive voltage from internal ADC
>100 mV.
Replace PC board.
121E
Operational PRV voltage reference test error:
PRV drive voltage from internal ADC >100 mV.
Replace PC board.
121F
Operational blower voltage reference test error:
blower drive voltage from MUX
>40 mV on second try, immediately following
first try.
Replace PC board.
122A
Corrupt resistance table: software error, invalid
data.
Replace PC board.
122B
Invalid mode: software error, selected mode
not supported in software.
Install PAV/T EPROM on PC board.
122C
Blower error: overcurrent, undervoltage,
shutdown.
1. Replace PC board.
2. Replace blower.
122D
Backup outlet pressure sensor error: data
cannot be read from sensor.
Replace PC board.
122E
Outlet difference error: excessive discrepancy
between outlet and backup outlet pressure
sensor readings
1. Replace PC board.
2. Replace air flow module (AFM).
3. Replace PSS.
122F
Backup outlet pressure sensor stuck: faulty
pressure sensor or ADC.
Replace PC board.
1220
Operational IFLR voltage reference error: drive
voltage read from internal ADC >40 MV on the
second try, immediately following first try.
Replace PC board.
Table 6-4: Troubleshooting
Vent Inop
Error Codes
Error code
Description
Corrective action