Chapter 7
Testing and Calibration
1062788
BiPAP Vision Service Manual © Respironics, Inc.
7-17
8. Press
MONITORING
, then select these settings in S/T mode:
9. Press
ALARMS
, then select these alarm limits:
10.
Monitoring screen parameters test:
Press
MONITORING
to go to the Monitoring screen, and verify the
following:
Rate
soft key indicator flashes when each timed breath is activated.
Each timed breath lasts approximately
1 s
.
IPAP
during a timed breath is 15 cmH
2
O.
EPAP
during a timed breath is 5 cmH
2
O.
Timed breath rate is 15 BPM (timed breath indicator display).
11.
Spontaneous breath test:
Occlude the test orifice for a few seconds. View the pressure
waveform, and create a small leak at the orifice to simulate a
spontaneous trigger (you might need to repeat this process a few
times).
Verify that the ventilator cycles to IPAP and that the Rate indicator
does not flash. Once a spontaneous breath is triggered, unblock the
test orifice.
12.
High pressure alarm test:
Press
ALARMS
to go to the Alarms screen, then set
Hi P
to 10
cmH
2
O. Wait for audible and visual alarms indicating a High Pressure
alarm. Return
Hi P
to 20 cmH
2
O and press Alarm reset to clear the
alarm.
13.
Low pressure alarm test:
Open the circuit to atmosphere for about 20 seconds to verify that the
Disconnect alarm activates after a few seconds. Press Alarm silence.
14. Verify that the Lo P alarm is activated. Occlude the ventilator outlet
and press Alarm reset to clear the alarm.
IPAP
= 15 cmH
2
O
Timed Insp
= 1.0 s
EPAP
= 5 cmH
2
O
IPAP Rise Time
= 0.1 s
Rate
= 15 BPM
Hi P
= 20 cmH
2
O
Lo MinVent
= Disabled
Lo P
= 10 cmH
2
O
Hi Rate
= 40 BPM
Lo P Delay
= 20 s
Lo Rate
= 10 BPM
Apnea
= Disabled