Testing and Calibration
7.8 Oxygen Module
(OM) Calibration
The OM calibration adjusts the zero voltage on the OM assembly main board.
Perform this calibration on any OM module that is suspected to be out of
tolerance, then perform the complete system final test on the ventilator.
NOTE:
Perform OM calibration on the existing OM before replacing any of its
components. If it is necessary to replace an OM component, test the OM
using the system final test (Section 7.7). If OM testing fails, perform OM
calibration then repeat the system final test.
NOTE:
If the OM still does not pass the system final test, follow standard repair
procedures.
Required equipment: Fluke 87 digital multimeter (DMM) or equivalent (3 1/2-
digit readout)
1. Remove the complete OM assembly except for its ground wire and
ribbon cable, and set next to the ventilator.
2. Connect the DMM to TP9 and TP13 (ground).
3. Power on the ventilator and allow it to complete its system self test.
Do
not
enter Monitoring.
4. Adjust R41 on the OM so that the DMM measures 0.225-0.235 VDC.
5. Reassemble the ventilator.