Troubleshooting
Table 6-4 describes
Vent Inop
error codes that apply to the MC, DC, and PC
boards.
Table 6-4: Troubleshooting
Vent Inop
Error Codes
Error code
Description
Corrective action
Vent Inop errors common to all subsystems:
Vent
Inop
with
Check
Vent
error
code 201
The ventilator continues operating when both a
Vent Inop
and a
Check Vent
(error code 201)
occur.
Perform service described in Section
6.7.
0
(MC, DC, PC)
Hardware error.
See error reported for a specific
subsystem.
601 (MC)
E01 (DC)
1601 (PC)
Microcontroller unit (MCU) detects spurious
interrupt.
Replace subsystem with lit error LED.
602 (MC)
E02 (DC)
1602 (PC)
MCU detects unassigned interrupt.
Replace subsystem with lit error LED.
603 (MC)
E03 (DC)
1603 (PC)
MCU detects bus interrupt.
Replace subsystem with lit error LED.
604 (MC)
E04 (DC)
1604 (PC)
MCU detects illegal instruction execution.
Replace subsystem with lit error LED.
605 (MC)
E05 (DC)
1605 (PC)
MCU detects breakpoint error.
Replace subsystem with lit error LED.
606 (MC)
E06 (DC)
1606 (PC)
MCU detects divide by zero error.
Replace subsystem with lit error LED.
607 (MC)
E07 (DC)
1607 (PC)
MCU detects municipalized interrupt.
Replace subsystem with lit error LED.
608 (MC)
E08 (DC)
1608 (PC)
MCU detects software interrupt execution.
Replace subsystem with lit error LED.
609 (MC)
E09 (DC)
1609 (PC)
MCU detects unused interrupt execution.
Replace subsystem with lit error LED.