Chapter 6
6-26
BiPAP Vision Service Manual © Respironics, Inc.
1062788
Troubleshooting
6.6 Interpreting Error
Codes
To interpret error codes, start by looking at all error codes and their definitions.
If one of the error codes indicates a hardware fault, this may be the “primary”
fault.
For example, if these error codes occur:
MC = 70E (Tack timeout)
PC = 0 (hardware failure)
DC = 11D1 (touch pad error)
In this example, the error code
11D1
indicates a touch pad error, and may be
the primary fault. This
Vent
Inop
error powers down the ventilator.
The error code
70E
indicates a timeout, because the MC is not receiving
communication from the PC.
The error code
0
indicates a hardware failure, because the ventilator is
powering down.