Chapter 4
4-4
BiPAP Vision Service Manual © Respironics, Inc.
1062788
Periodic Maintenance
4.4 Oxygen Regulator
Filter
A dirty oxygen regulator filter can affect ventilator performance. Examine the
filter for cleanliness and damage before each use and as required during
operation. Replace the oxygen regulator filter (Figure 4-3, Figure 4-4) as
required. See Chapter 5 for component removal/installation instructions.
•
If needed, clean the regulator bowl with mild soap and water, then
rinse and dry completely.
Figure 4-3: Removing the Oxygen Regulator Bowl and Filter
(OM s/n 299999 and below)
Oxygen module s/n here
Oxygen module regulator bowl
(P/N 582154)
Oxygen module regulator filter
(P/N 1007547)